BROADBAND TERMINATED CURRENT TRANSFORMER DATA

These current transformers are designed for use as general laboratory instruments
where band width and high-frequency response are to be combined with ease of use

 
Models 13D, 13G, 13W
Models 13D, 13G, 13W

Model_13
TYPICAL VALUES Models 13D, 13G, 13W

Model
number

Output
V/A(1)

RECTANGULAR PULSE

SINEWAVE

DC saturation
Current
Isat
A

Max. rms
current
Irms
A

Peak
current
Ipeak
A

Price
GBP
(3)

Droop
%/ms

Rise
time
ns

I-t Capability

3db Bandwidth

Ipeak/f
A/Hz

No
bias
mA s

With
bias
mA s(2)

If
Hz

hf
MHz

13D0020

0.020

1.2

140

1800

3200

2

3

12

18

200

25000
710

13D0050

0.050

3

40

800

1500

5

10

5

18

100

10000
658

13G0050

0.050

0.25

40

1200

2200

0.5

10

8

3

100

10000
645

13W0050

0.050

0.04

40

500

900

0.1

10

3

0.15

100

10000
676

13D0100

0.10

10

16

250

450

15

25

1.7

18

60

5000
572

13G0100

0.10

1

16

400

700

1.6

25

2.5

3

60

5000
559

13W0100

0.10

0.15

16

150

250

0.3

25

0.9

0.15

60

5000
588

13D0200

0.20

30

14

70

130

60

30

0.4

18

33

2500
534

13G0200

0.20

4

14

100

200

10

30

0.7

3

33

2500
515

13W0200

0.20

0.6

14

40

70

1

30

0.25

0.15

33

2500
552

13D0500

0.50

170

10

13

25

220

40

0.08

18

14

1000
485

13G0500

0.50

20

10

15

30

35

40

0.12

3

14

1000
469

13W0500

0.50

3

10

7

13

5

40

0.04

0.15

14

1000
498

13D1000

1.0

700

7

3

5

1000

60

0.02

18

7

500
431

13G1000

1.0

70

7

4

8

130

60

0.03

3

7

500
418

13Wl000

1.0

12

7

1.5

3

30

60

0.01

0.15

7

500
442
13W5000
5.0
240
4
0.06
0.11
400
100
0.0004
0.15
2
25
431

 

Models 58D, 58M
Models 58D, 58M

Model_58-line_drawing

TYPICAL VALUES Models 58D, 58M
Model
number
Output
V/A(1)
RECTANGULAR PULSE
SINEWAVE
DC saturation
current
Isat
A
Max. rms
current
 Irms
A
Peak
current
Ipeak
 A

Price
GBP
(3)

Droop
%/ms
Rise
time
ns
I-t Capability
3dB Bandwidth
Ipeak/f

A/Hz
No bias
mA s
With
bias
mA s
lf
Hz
hf
MHz
58DH010
0.010
0.6
80
10000
19000
1
5
60
60
400
50000
990
58MH010
0.010
0.025
80
13000
24000
0.04
5
85
2.0
400
50000
810
58D0020
0.020
0.6
200
10000
18000
1
2
60
60
500
20000
930
58M0020
0.020
0.04
200
12000
20000
0.06
2
70
2
500
20000
810
58D0050
0.050
2.5
80
2500
4900
4.0
5
15
60
200
10000
930
58M0050
0.050
0.1
80
3400
6000
0.15
5
20
2.0
200
10000
785
58D0100
0.10
9
40
800
1400
15
10
4.0
60
100
5000
935
58DH100
0.10
30
20
250
450
50
20
1.5
60
100
5000
935
58M0100
0.10
0.3
40
1000
1800
0.5
10
6.0
2.0
100
5000
715
58MH100
0.10
1
20
300
500
1.5
20
2.0
2.0
100
5000
715
58D0200
0.20
60
20
120
230
90
20
0.8
60
50
2500
750
58M0200
0.20
2
20
150
280
4
20
1
2.0
50
2500
700
58M0500
0.50
9
20
40
70
15
20
0.2
2.0
20
1000
680
58M1000
1.0
40
20
10
18
55
20
0.06
2.0
12
500
640
58M5000
5.0
800
14
0.4
0.7
1500
30
0.002
2.0
2
100
600
NOTES:

Source impedance - All models have a source impedance of 50R
Load impedance - In the above tables this is assumed to be >50k
50R load - With a 50R load I-t capability is increased by between 20% and 80% and droop and lf cut-off point are reduced by a factor of 0.6 to 0.85. A 50R load at the receiver and a 50R line should be used where long cables are employed or where the highest frequency response is required.
Output – Accuracy +/-0.5% at mid-log frequency point, +/-2% within one decade of 3dB cut-off points. The use of a 50R load will halve the sensitivity
I-t No bias – Lilco broadband current transformers do not require the application any biasing current to achieve this level of I-t. The use of a 50R load will enhance the I-t capability significantly.
I-t with bias – these levels are achieved with a dc off-set current equal to Isat. The use of a 50R load will enhance the I-t capability further
Price – small quantity price, 1-2 off
58M0020 and 58D0020 – these broadband current transformers have been introduced to satisfy the measurement demands of EN61000-4-11

TOP OF PAGE
RETURN TO HOME PAGE